Using the RamanStation 400 .
99
Figure 32 System verification/calibration flow diagram
If wavelength calibration verification fails and there is an error message stating that peak
wavelengths are outside limits, then wavelength and laser frequency calibrations should be
performed. When these calibrations have finished, repeat the wavelength verification to
ensure that the instrument has been brought back within specification.
If the wavelength calibration verification fails because the spectral intensity is low, then this
may also be due to the calibration sample holder not being placed in the sample accessory,
or that it is not correctly positioned, optimally focused or because the laser key is not
switched on. Then repeat the wavelength calibration verification to ensure that instrument
has been brought back within specification. If the instrument fails the wavelength calibration
verification again, please contact PerkinElmer.
Summary of Contents for RamanStation 400
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanStation 400 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 28: ...28 RamanStation 400 Series Getting Started Guide...
Page 29: ...Raman History Theory and Instrumentation...
Page 37: ...Preparing Your Facility...
Page 42: ...42 RamanStation 400 Series Getting Started Guide...
Page 43: ...Installing Your RamanStation 400...
Page 71: ...Introduction to the RamanStation 400...
Page 76: ...76 RamanStation 400 Series Getting Started Guide...
Page 77: ...Fiber Optic Probe...
Page 84: ...84 RamanStation 400 Series Getting Started Guide...
Page 85: ...Using the RamanStation 400...
Page 106: ...106 RamanStation 400 Series Getting Started Guide...
Page 107: ...Maintenance and Troubleshooting...
Page 118: ...118 RamanStation 400 Series Getting Started Guide...