100 . RamanStation 400 Series Getting Started Guide
SOP-10: Wavelength Calibration Verification of the
RamanStation 400
1.
Ensure the instrument is switched on by following SOP-1.
2.
Open the Scan and Instrument Setup dialog, and on the Hardware tab ensure that the
laser power is set to 100%.
3.
Focus on a sample of polystyrene using the autofocus buttons and Spectral Preview.
4.
When the polystyrene sample is optimally focused, close the Scan and Instrument Setup
dialog by clicking Exit.
5.
Select Validate
and then Instrument from the Instrument menu in Spectrum.
The Validation / Calibration dialog is displayed.
6.
Select the Verification tab.
Figure 33 The Validation / Calibration dialog
7.
Select Wavelength Calibration Verification, and then click OK.
The wavelength calibration verification routine begins. This will take a few minutes.
After completing the wavelength calibration verification routine, an audit log is
generated.
8.
Remove the polystyrene sample and store safely.
The instrument is now ready for use.
In the unlikely event that the wavelength calibration verification should fail, follow Figure 32.
If the RamanStation 400 wavelength calibration verification fails for the sample
compartment, this does not necessarily mean that the performance of the fiber optic probe is
also affected.
Further details on the wavelength calibration verification are provided as a spectral (*.sp)
file, which is saved in
C:\pel_data\config\Raman\Calibration Files\Logs
folder.
Summary of Contents for RamanStation 400
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY...
Page 5: ...Introduction...
Page 10: ...10 RamanStation 400 Series Getting Started Guide...
Page 11: ...Safety Practices...
Page 28: ...28 RamanStation 400 Series Getting Started Guide...
Page 29: ...Raman History Theory and Instrumentation...
Page 37: ...Preparing Your Facility...
Page 42: ...42 RamanStation 400 Series Getting Started Guide...
Page 43: ...Installing Your RamanStation 400...
Page 71: ...Introduction to the RamanStation 400...
Page 76: ...76 RamanStation 400 Series Getting Started Guide...
Page 77: ...Fiber Optic Probe...
Page 84: ...84 RamanStation 400 Series Getting Started Guide...
Page 85: ...Using the RamanStation 400...
Page 106: ...106 RamanStation 400 Series Getting Started Guide...
Page 107: ...Maintenance and Troubleshooting...
Page 118: ...118 RamanStation 400 Series Getting Started Guide...