Fiber Optic Probe . 63
9.
If you have a standard probe, connect the DIN connector to the nearby socket marked
LED FIBER #1.
OR
If you have a triggered probe, connect the USB extender cable to a USB port on your
PC.
10.
Reconnect the power cables.
Removing the FOP
If it is necessary to remove the FOP, for example, for replacement or transportation, please
follow this procedure carefully:
1.
Ensure the attenuator cap is attached to the FOP.
2.
Switch off the instrument and remove the main power cable and, if appropriate, the
stage controller power cable.
3.
Carefully move the instrument to permit easier access to the rear panel.
4.
If you have a standard probe, remove the DIN connector from the socket marked
LED FIBER #1.
OR
If you have a triggered probe, disconnect the connect the USB extender cable from the
USB port on your PC.
5.
Carefully remove the retained ends of the FOP by unscrewing and pulling the fiber optic
cables away from the EXC and COL couplers perpendicularly.
Be extremely careful when removing the FOP as you can easily damage the instrument
and the fiber optic cables.
6.
Replace the protective plastic caps on the EXC and COL ends of the FOP that have been
removed from the rear panel.
7.
Fasten the retained screw-in caps onto the couplers on the rear panel of the instrument.
8.
Replace the fiber port cover assembly if required (see
Attaching the Fiber Port Cover
Assembly
on page 46).
9.
Reconnect the power cables.
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...