Using the RamanStation 400 . 69
General SOPs
SOP-1: Turning the Instrument On
1.
Switch on the PC and monitor.
2.
Login to Windows.
3.
Ensure the laser power key is in the ARM position on the rear panel of the instrument.
4.
Ensure the main power supply is connected to a mains outlet and that the outlet is
switched on.
5.
To turn on the spectrometer, press the PSU power switch ON (Figure 32):
- ON
Ο
- OFF
Figure 32 The PSU power switch
6.
Check that the Power LED on the instrument keypad is On.
The Laser LED and the Active LED should be off. If the Power LED is not on, please turn
the instrument off at the PSU and check all of the connections, any power socket
switches and then switch the PSU on again.
7.
If appropriate, connect the stage controller box to a mains outlet and ensure that the
outlet and the controller are switched on.
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...