SOP-11: Laser Frequency Calibration
Laser frequency calibration should only be performed if the instrument has failed a
wavelength calibration verification.
1.
Ensure the RamanStation 400 is switched on.
If not, follow SOP-1.
2.
Select
Raman Instrument
from the Setup menu.
The Setup Instrument tabs are displayed
3.
Ensure that
Sample Area XYZ Stage
or
Sample Area
, as applicable, is selected as the
Accessory on the Setup Instrument Basic tab.
4.
Ensure that the lens snout is set to vertical physically, and that the
Vertical
orientation
is selected in the software.
5.
Load a polystyrene sample (L1181228) into a suitable sample holder.
6.
Select the appropriate
Sample Holder
in the drop down list and load the sample holder
onto the sample stage, following the instructions on-screen if you have a motorized
stage.
7.
Ensure that the RamanStation 400 Series sample compartment is closed and, if you
have a RamanStation 400F, ensure that any external accessory is capped or covered, to
prevent stray light interfering with the measurement.
If you have a RamanMicro 300 Accessory connected to your RamanStation 400F, ensure
that the microscope illumination is switched off.
8.
Select the Setup Instrument Advanced tab and ensure that the laser power is set to
100%.
9.
Focus on the polystyrene sample.
If you have a motorized stage, click
Raman AutoFocus
on the Setup Sample Area
XYZ Stage tab to focus using the strongest Raman signal. The Focus Limits should be
set to
Full Range
.
10.
When the polystyrene sample is optimally focused for Raman signal, select the Setup
Instrument Advanced tab and click
Toolbox
.
The Adjustments Toolbox dialog is displayed.
11.
Select
.
The Frequency Calibration dialog is displayed.
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...