Installing Your RamanStation 400 Series . 51
Figure 22 Triggered Fiber Optic Probe Holster
1.
Make sure that the instrument is switched off and disconnected from the mains supply.
2.
Carefully move the instrument so that the right-hand side is positioned over the edge of
your workbench.
WARNING
While the instrument is not stood flat on the bench, you must keep it
supported at all times.
3.
Locate and remove the two screws securing the holster bracket to the base of the
instrument (located behind the front foot).
Figure 23 Base of the RamanStation 400 Series instrument
4.
Return the instrument to its correct position on your workbench.
Screws
Right front foot of
RamanStation