94 . RamanStation 400 Series Getting Started Guide
Replacing the Stage Controller Fuse
NOTE: This section is only applicable if you have an motorized XYZ stage.
WARNING
Electrical hazard
To prevent potential injury to yourself and damage to the instrument,
switch OFF all components in the system and disconnect them from the
line power supply before you alter, or make any new, electrical
connections.
Removal of the protective housing or exterior panels on any instrument
will allow potentially hazardous Class 3B laser radiation to be emitted
and will invalidate the instrument warranty.
1.
Switch off the stage controller, disconnect it from the power supply and remove the
mains cord.
2.
Insert a screwdriver into the slot at the side of the fuse drawer, and pull out and flip to
one side over the mains inlet.
The fuse may now be removed.
3.
Fit the replacement fuse into the fuse drawer.
Make sure that you fit the fuse in the top slot. You require a 1.6 A time-lag fuse
(09991641). Always use the recommended replacement fuse.
4.
Refit the fuse drawer.
The only other fuses that you may have to replace are those in the plugs of the mains power
cords for the spectrometer and microscope (UK-type plugs only).
Summary of Contents for RamanStation 400 Series
Page 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 RamanStation 400 Series Getting Started Guide ...
Page 11: ...Safety Practices ...
Page 27: ...Raman History Theory and Instrumentation ...
Page 34: ...34 RamanStation 400 Series Getting Started Guide ...
Page 35: ...Preparing Your Facility ...
Page 39: ...Installing Your RamanStation 400 Series ...
Page 52: ...52 RamanStation 400 Series Getting Started Guide ...
Page 53: ...Introduction to the RamanStation 400 Series ...
Page 59: ...Fiber Optic Probe ...
Page 66: ...66 RamanStation 400 Series Getting Started Guide ...
Page 67: ...Using the RamanStation 400 ...