Group
Parameter
Description
Size
Measurement
Accumulation
Times
Enter the number of integration cycles for the
measurement; the default is 70. At this value, the
software determines 70 ACFs and calculates 70
realtime diameters from each ACF, ultimately
calculating one averaged ACF from 70 ACFs to give
the mean diameter.
Increase the number of cycles if the intensity of the
sample is weak or if the signal is too noisy to obtain a
stable autocorrelation function. If the intensity is less
than 10,000 cps, use the following formula as a
guideline to input the number of cycles:
Number of integration cycles = 10,000 (cps) /
intensity of the sample (cps) x 10
Size
Measurement
(Details)
This group of
parameters
displays when
Linear
correlator is
selected.
Correlation Channel
Select the correlation channel of the correlation
calculation. The correlation channel represents the
number of bins into which the size data are
distributed. It is the size resolution (the more
channels, the more divisions on the size scale, i.e., X-
axis).
Select
Auto
if you are not sure which channel to use.
Correlation Method
Select
TD
(Time Domain),
TI
(Time Interval), or
Auto
. Typically,
TD
is for large particles having
stronger scattering levels, and
TI
for small particles
of less than 10 nm having weaker scattering levels.
Select
Auto
if you are not sure which to choose.
Sampling Time (
s)
Select the sampling time from the list. The sampling
time is the time after which the signal is collected
from the sample by the instrument. Appropriate
sampling time must be selected to get a smoother
ACF and for accurate size measurement.
Select
Auto
if you aren’t sure which time to use.
Titrator
Titration Mode
This parameter determines the mode for pH titration
vs. size, diluting/priming, or circulation of the
sample.
Choose
pH titration
,
Pipetting
, or
Circulation
.
The Titration Mode is applicable only when you are
using a flow cell.
SOP Designer
NanoPlus Operator’s Manual
3-28
Mar 2014
Summary of Contents for NanoPlus
Page 1: ...NanoPlus Zeta Potential Nano Particle Analyzer Operator s Manual V5 22 P09 42801 01 March 2014...
Page 80: ......
Page 114: ......
Page 176: ...Consumables NanoPlus Operator s Manual A 30 Mar 2014 Tubing...
Page 188: ......
Page 192: ......
Page 213: ...NanoPlus Operator s Manual Temperature Gradient Plot Mar 2014 D 21 With Temperature Marker...
Page 232: ......
Page 246: ......
Page 252: ......