SOP Designer
NanoPlus Operator’s Manual
3-26
Mar 2014
Common Parameters
The following table lists the common cell parameters for
Size
and
Zeta
Potential
analyses.
Parameters are listed in alphabetical order;
Advanced
ones are
italicized
for easy recognition.
Group
Parameter
Description
General
Cell Center X (mm)
Cell Center Z (mm)
Displays the optimal positions of the cell on the X and
Z axes. The values are the default ones unless a cell
center adjustment has been performed and saved, in
which case it will be shown. These fields are not
shown for Cell Types 2 and 4.
Detecting the Optimal Cell Position,
additional information.
Cell Name
Select the field to display a button to the right. Click
the button to display a list of registered cells for the
Measurement type you choose (Size or Zeta). This
example shows the list for Size.
Be sure the sample conductivity is
lower than ~3 ms/cm when using the
High Concentration cell. A higher
conductivity may damage the cell.
If performing gradient temperature analyses, you
must use one of the following cells; analysis will not
proceed with any other type.
Measurement
Cell Type
Temperature
Range (
C)
Size
Size cell (glass)
5.0 to 90.0
Size cell (Dispo)
5.0 to 50.0
Size cell (micro)
5.0 to 90.0
Zeta
Flow cell
5.0 to 90.0
Zeta cell (Dispo)
5.0 to 50.0
Summary of Contents for NanoPlus
Page 1: ...NanoPlus Zeta Potential Nano Particle Analyzer Operator s Manual V5 22 P09 42801 01 March 2014...
Page 80: ......
Page 114: ......
Page 176: ...Consumables NanoPlus Operator s Manual A 30 Mar 2014 Tubing...
Page 188: ......
Page 192: ......
Page 213: ...NanoPlus Operator s Manual Temperature Gradient Plot Mar 2014 D 21 With Temperature Marker...
Page 232: ......
Page 246: ......
Page 252: ......