Maintenance
NanoPlus Operator’s Manual
5-8
Mar 2014
Flow Cell
This section describes how to disassemble and assemble the Flow Cell, as well as how to wash the
glass cell.
1
Cell Holder
6
Cell Guide (2)
2
Clamping Knob (2)
7
Luer Fitting (2)
3
O-ring (8); (6 spares)
8
Glass Cell
4
Electrode (2)
9
Cell Stopper
5
Sealing Rubber (2); (6 spares)
10
Plug (2)
1
2
3
4
5
6
7
8
9
10
Assembled
Disassembled
Disassembling the Flow Cell
Wear protective gloves when disassembling the cell. There may be residual acidic/
alkaline sample solution remaining in the cell.
1.
Drain the sample solution in the cell.
2.
Remove the cell stopper by turning it clockwise.
3.
Remove the Luer fitting by turning it counterclockwise.