NanoPlus Operator’s Manual
SOP Designer
Mar 2014
3-23
Zeta Parameters
Group
Parameter
Description
General
Comment
Enables you to enter a comment about the analysis
conditions. You can enter up to 40 characters.
Condition name
Identifies the analysis condition. Use a unique, easily
recognizable name. You can use up to 40 characters.
Conversion Equation
This is used to select the equation for calculating the
zeta potential from the mobility. You may select
Smoluchowski
,
Huckel
, or
Other
. Generally,
Smoluchowski
is used for aqueous samples and
Huckel
is used for organic samples.
You must enter a coefficient when you select
Other
.
This coefficient is substituted for the “k” in the
following equation:
Z
k
-------U
=
where
Z = Zeta potential
= Viscosity of the solution
= Dielectric constant of the solution
U = Mobility