NanoPlus Operator’s Manual
Maintenance
Mar 2014
5-13
5.
Remove the transparent electrode holder.
Knob B
Electrode clamp
Transparent electrode holder
6.
Pull the electrode rod from the transparent electrode holder, and remove the transparent
electrode.
Surface of electrode
Optical surface
Transparent electrode holder
Electrode rod
Transparent electrode
7.
Remove the heat conductive adapter upward, and remove the high concentration cell.
Base
High concentration cell
Heat conductive adapter
8.
Remove the large O-ring from the high concentration cell.
Teflon seal
O-ring (large)
9.
Rinse the high concentration cell.