Zeta
Measurement
(
continued
)
Equilibration (sec)
Allows you to enter a delay time for the measurement
to begin. You can enter up to 100,000 seconds. The
equilibration time starts when the run begins; the
measurement starts when the equilibration time has
elapsed.
Repetition
This is the number of times to repeat the
measurement. Maximum of 10,000 repetitions.
Wait Time (sec)
Enter the time (in seconds) to wait between repeated
measurements. The range is 0 - 100,000 sec.
Others
Pinhole (
m)
Sets the pinhole size. The default is 50
m and
typically is the one used. The available sizes are 20,
50, and 100 selected from the drop-down list.
If the pinhole setting is set at 100,
the intensity (scattering strength)
that can be obtained is increased.
However, the state of the coherence
deteriorates, reducing the signal-to-
noise ratio.
NanoPlus Operator’s Manual
SOP Designer
Mar 2014
3-17