17.8 Level meter board
Helium probe specifications
Parameter
Specification
Excitation current settings
20 mA to 280 mA in 20 mA steps
Excitation current accuracy
±5%
Excitation voltage
40 V maximum
Probe resistance
0 to 300 ohms
Probe resistance measurement accuracy
±0.2%
Probe resistance measurement stability
±0.1%
Probe resistance ADC measurement resolution
Better than 0.005%
Probe connections isolation to chassis
>50 VDC
Nitrogen probe specifications
Parameter
Specification
Probe source voltage
12 VDC
Probe source oscillator frequency range
5 kHz to 90 kHz
Probe sensing resolution (level)
0.1%
Probe sensing accuracy (level)
0.5%
17.9 Pressure board
Parameter
Specification
Measurement absolute accuracy
±0.1% full scale in each range
Measurement Resolution
24 bits
Voltage Measurement Ranges
10V, 1.6V, 0.8V, 0.4V, 0.2V, 0-0.1V (Auto-ranging)
Current Measurement Ranges
4-20mA (20mA, 16mA, 8mA, 4mA Auto-ranging)
Measurement Frequency
1Hz-10Hz
Sensor Excitation Voltage
0 V(off) , 10V±5% or 15V±5%
Sensor Excitation Current
0 - 20 mA
Sensor Excitation Current Trip
>30mA
Short Circuit Protection Duration
Indefinite
Isolation From Mercury Chassis
50 Volts
Over Voltage Protection
Diode on each connector pin: 50V/200mA