Operator's Manual
Issue 14
/
Mar 2016 / UMC0071
Mercury
iPS
©2016 Oxford Instruments NanoScience. All rights reserved.
Page
154
Calibration error
Recalibrate
Firmware error
Error in Board firmware. Update firmware.
Board not configured
Firmware not loaded correctly. Update firmware.
Auxiliary I/O board
Firmware error
Error in Board firmware. Update firmware.
Board not configured
Firmware not loaded correctly. Update firmware.
Level meter board
Open circuit
Look for open circuit on the probe input
Short circuit
Look for short circuit on the probe input
ADC error
This is on-board diagnostic. Recalibrate.
Over demand
This is on-board diagnostic. Recalibrate.
Over temperature
Firmware error
Error in board firmware. Re-start iPS
Board not configured
Firmware not loaded correctly. Update firmware.
No reserve
Autofill stops after 2hr The auto fill valve is open but the dewar is not filling
Pressure board
Open circuit
Look for open circuit on the sensor input
Short circuit
Look for short circuit on the sensor input
Calibration error
Recalibrate
Firmware error
Error in board firmware Update firmware.
Board not configured
Firmware not loaded correctly. Update firmware.
Over current
Look for partial short circuits (low impedance)
Current leakage
Look for short to ground
Power on fail
Re-start iPS
Checksum fail
Re-start iPS
Clock fail
Re-start iPS
ADC fail
On-board diagnostic. Recalibrate.
Mains fail
Re-start iPS
Reference fail
Re-start iPS
12V fail
Re-start iPS
-12V fail
Re-start iPS
8V fail
Re-start iPS
-8V fail
Re-start iPS
Amp gain error
Re-start iPS
Amp offset error
Re-start iPS
ADC offset error
Re-start iPS
ADC PGA error
Re-start iPS
ADC XTAL failure
Re-start iPS
Exci error
Re-start iPS
Excitation - error
Re-start iPS
If you are experiencing difficulties, please
check the relevant sections of this manual