11
1.1.2 High voltage hazard
HIGH VOLTAGE
Danger of death or serious injury. Switch off the external AC
electrical supply and disconnect and remove the external supply
cable, before maintaining this equipment or removing any covers.
HAUTE TENSION
Danger de mort ou de blessure grave. Sectionner l'alimentation
électrique C.A. externe puis déconnecter et retirer le cordon
d'alimentation externe avant d'effectuer toute maintenance de cet
équipement ou d'en retirer les capots.
This equipment contains high voltages. Isolate this equipment before removing the screws and
covers. To isolate this equipment, switch off the external AC electrical supply, and disconnect
and remove the external supply cable. Do not restrict access to the AC electrical supply at any
time. The AC electrical supply cable must remain accessible for disconnection of the equipment.
1.1.3 Maintenance and adjustment
Only qualified and authorised persons must service or repair this equipment.
Under no circumstances must the user adjust or repair this equipment while the electrical power
supply is connected.
Ensure that the equipment is disconnected from the AC electrical power supply before the
covers are removed. It is NOT sufficient to turn off the unit from the power-down button on the
front panel. It is also NOT sufficient to turn off the mains power switch
on the rear
panel. The unit MUST be disconnected from the mains supply. Also, the equipment contains
capacitors that may remain charged after AC electrical power is removed. Wait at least 15
minutes after disconnecting the mains supply to ensure all the capacitors are discharged before
starting work.
1.1.4 Restrictions on use
The equipment is not suitable for use in explosive, flammable or hazardous environments.
The equipment does not provide protection against the ingress of water. The equipment must
be positioned so that it will not be exposed to water ingress.