Operator's Manual
Issue 14
/
Mar 2016 / UMC0071
Mercury
iPS
©2016 Oxford Instruments NanoScience. All rights reserved.
Page
121
R2
Output Current
A
R3
Unused
R4
Output Current
A
R5
Set point (Target current)
A
R6
Target current sweep rate
A/min
R7
“Output” Field
T
R8
Set point (Target field)
T
R9
Target field sweep rate
T/min
R10
Temperature**
K
R11
Liquid Nitrogen level†
%
R12
Liquid Helium level†
%
R13
Unused
R14
Output Current
A
R15
Voltage limit
V
R16
Magnet current
A
R17
Unused
R18
Magnet field strength
T
R19
Unused
R20
Switch heater current
mA
R21
Negative current limit
A
R22
Positive current limit
A
R23
Unused
R24
Magnet Inductance
H
Note: ** The temperature measured by the MB1.T1 sensor input channel. This will typically
be the magnet temperature or lambda plate temperature.
† Assuming a cryogen level meter card is fitted in the iPS-Master.
10.4.2.2 V command - Read Version
The VERSION command requires no parameters. It returns a message indicating the
instrument type and firmware version number.
10.4.2.3 X command - Examine Status
The EXAMINE command allows the computer to read the present iPS STATUS. The command
requires no parameters and returns a message string with a fixed length of 13 characters.
The returned string is of the form
XmnAnCnHnMmn
where