SECTION 7 - BASIC TROUBLESHOOTING, ELECTRICAL INFORMATION, & SCHEMATICS
3121827
– JLG Lift –
7-19
7.3 MULTIMETER BASICS
A wide variety of multimeters or Volt Ohm Meters (VOM)
can be used for troubleshooting your equipment. This
section shows diagrams of a common, digital VOM con-
figured for several different circuit measurements. Instruc-
tions for your VOM may vary. Please consult the meter
operator’s manual for more information.
Grounding
"Grounding the meter" means to take the black lead
(which is connected to the COM (common) or negative
port) and touch it to a good path to the negative side of
the Voltage source.
Backprobing
To "backprobe" means to take the measurement by
accessing a connector’s contact on the same side as the
wires, the back of the connector. Readings can be done
while maintaining circuit continuity this way. If the connec-
tor is the sealed type, great care must be taken to avoid
damaging the seal around the wire. It is best to use
probes or probe tips specifically designed for this tech-
nique, especially on sealed connectors. Whenever possi-
ble insert probes into the side of the connector such that
the test also checks both terminals of the connection. It is
possible to inspect a connection within a closed connec-
tor by backprobing both sides of a connector terminal and
measuring resistance. Do this after giving each wire a
gentle pull to ensure the wires are still attached to the con-
tact and contacts are seated in the connector.
Min/Max
Use of the "Min/Max" recording feature of some meters
can help when taking measurements of intermittent condi-
tions while alone. For example, you can read the Voltage
applied to a solenoid when it is only operational while a
switch, far from the solenoid and meter, is held down.
Polarity
Getting a negative Voltage or current reading when
expecting a positive reading frequently means the leads
are reversed. Check what reading is expected, the loca-
tion of the signal and that the leads are connected to the
device under test correctly. Also check that the lead on the
"COM" port goes to the Ground or negative side of the sig-
nal and the lead on the other port goes to the positive side
of the signal.
Scale
M = Mega = 1,000,000 * (Displayed Number)
k = kilo = 1,000 * (Displayed Number)
m = milli = (Displayed Number) / 1,000
μ
= micro = (Displayed Number) / 1,000,000
Example: 1.2 k
= 1200
Example: 50 mA = 0.05 A
Voltage Measurement
• If meter is not auto ranging, set it to the correct
range (See multimeter’s operation manual)
• Use firm contact with meter leads
Figure 7-1. Voltage Measurement (DC)
Summary of Contents for JLG E400AJP
Page 2: ......
Page 32: ...SECTION 1 SPECIFICATIONS 1 18 JLG Lift 3121827 NOTES...
Page 42: ...SECTION 2 GENERAL 2 10 JLG Lift 3121827 NOTES...
Page 54: ...SECTION 3 CHASSIS TURNTABLE 3 12 JLG Lift 3121827 Figure 3 3 Speed Sensor Orientation...
Page 60: ...SECTION 3 CHASSIS TURNTABLE 3 18 JLG Lift 3121827 Figure 3 7 Steering Components and Spindles...
Page 62: ...SECTION 3 CHASSIS TURNTABLE 3 20 JLG Lift 3121827 Figure 3 9 Tilt Sensor Location...
Page 86: ...SECTION 3 CHASSIS TURNTABLE 3 44 JLG Lift 3121827 Figure 3 16 Swing Components...
Page 88: ...SECTION 3 CHASSIS TURNTABLE 3 46 JLG Lift 3121827 Figure 3 18 Battery Cable Connections...
Page 90: ...SECTION 3 CHASSIS TURNTABLE 3 48 JLG Lift 3121827 Figure 3 20 On Board Generator...
Page 97: ...SECTION 3 CHASSIS TURNTABLE 3121827 JLG Lift 3 55 Figure 3 22 Generator Components...
Page 116: ...SECTION 3 CHASSIS TURNTABLE 3 74 JLG Lift 3121827 NOTES...
Page 127: ...SECTION 4 BOOM PLATFORM 3121827 JLG Lift 4 11 Figure 4 10 Boom Limit Switches...
Page 140: ...SECTION 4 BOOM PLATFORM 4 24 JLG Lift 3121827 Figure 4 13 Rotator Counterbalance Valve...
Page 178: ...SECTION 5 HYDRAULICS 5 24 JLG Lift 3121827 Figure 5 26 HydraForce Cartridge Torque Value Chart...
Page 214: ...SECTION 6 JLG CONTROL SYSTEM 6 34 JLG Lift 3121827 NOTES...
Page 257: ......