WWW.NNC.IR
Error Messages and Troubleshooting
828
Vision System FH/FZ5 Series
User’s Manual (Z340)
The problem occurred in
the system.
This is displayed when the significant abnormality occurs in the controller system. Please
contact one of our branches or regional offices.
The problem occurred in
system date. The system
battery is insufficient.
Low battery (for the backup of date and time data) level. The batteries need to be replaced.
Please return old batteries to one of our branches or regional offices.
Failed to switch scene.
Is a USB disk set to the controller?
A USB disk is needed to read scenes after scene group 1.
Failed to switch scene
group or save scene group
on switch.
The cause of the failure to switch or to save when switching may be that the USBDisk was
removed from the controller. Set a USBDisk to the controller and try again.
Failed to clear scene
group.
Confirm that a USBDisk is set to the controller and try again.
Failed to load scene group.
Data is corrupted, or
memory is insufficient.
Scene group data starts
with initialized status.
The following causes are suspected:
• The power may have been cut off during the last data save.
• Because the operation mode was switched, the required memory size may have increased
and memory became insufficient.
The camera connected is
not the same as the one
used for the last save.
Please check.
Check if the camera is connected correctly.
This message is displayed when the system and the scene group 0 data edited using the
simulation software are loaded and the unit is restarted.
Save the setting to the controller and then restart.
Cannot read selected file.
Confirm selected file once
again.
Check the following points.
• Have you removed the USBDisk from the controller after selecting files on the USBDisk?
• Have you deleted the selected file? (Such as the case when you selected a file on the
RAMDisk and the file was deleted via FTP.)
In addition, the selected file may be corrupt.
The communication time-
out is occurred.
Switch off controller, verify the following contents and then restart.
• Is cable connected correctly?
• Does it comply with communication specifications of external devices?
• Are external devices functioning normally?
If error is not resolved after confirmation, the controller may be damaged. Please discuss this
with one of our branches or regional offices.
Reference:
Non-procedure Communications - Communications Specifications Settings
in
the
Vision System FH/FZ5 Series User’s Manual for Communications Settings
(Cat No. Z342)
Failed to transfer data.
The free capacity of
RAMDisk may insufficient.
Increase the free capacity
of RAMDisk and then
perform this operation
again.
Clean up the content of RAMDisk to increase free space. If this error occurs even though
enough free space is available, scene group data may be too large. Review the inspection flow.
Error in input range. Please
input using the correct
range.
Please verify range for setting and set up again.
File name contains invalid
character.
Please confirm that characters such as \ / , : ; * ? " < > |& . SPC (space) are not included in the
file name.
Failed to save file. Please
check.
Check the following points.
• Is the memory of the save destination sufficient?
• For cases where the save destination is the folder in the USBDisk, is the USBDisk connected
to the controller? Or, the controller may not have detected the USBDisk.
Error message
Troubleshooting
Summary of Contents for fh-1 series
Page 8: ...WWW NNC IR 6 Vision System FH FZ5 Series User s Manual Z340...
Page 14: ...WWW NNC IR 12 Vision System FH FZ5 Series User s Manual Z340 MEMO...
Page 22: ...WWW NNC IR 20 Vision System FH FZ5 Series User s Manual Z340 MEMO...
Page 48: ...WWW NNC IR Initializing the Controller 46 Vision System FH FZ5 Series User s Manual Z340 MEMO...
Page 68: ...WWW NNC IR Editing Scene Groups 66 Vision System FH FZ5 Series User s Manual Z340 MEMO...
Page 881: ...WWW NNC IR...