OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
41
Maintenance
Changing a Fuse
If the SGSM Electronics Box does not power on, check the fuse.
1.
Unplug the power cord leading to the SGSM Electronics Box.
2. Remove the fuse holder next to the power receptacle.
3. If either fuse is blown, replace it with a new one.
The SGSM Electronics Box uses two 2 amp fuses.
4. Put the fuse holder back into the slot.
5. Plug the power cord back in.
6. Turn on the Electronics Box to ensure it is working again.
ALWAYS connect the SGSM head before powering up the electronics
box. NEVER plug in the SGSM head when the box is already on. This
could damage the SGSM board.
Power Switch
Power Receptacle
Fuse Holder