6. Install the two bushings. One bushing goes into the hole in the top of the
top cell cap. The other goes inside the Mounting Adapter.
Bushing
(#120-53-42)
Bushing
(#120-53-42)
Backup Ring
(#120-53-58)
O-ring
(#120-53-52)
Mounting
Adapter
Retainer Ring (#120-53-82)
High Pressure SGSM Only
7. Place the upper backup ring (#120-53-58) around the top of the Mounting
Adapter with the o-ring (#120-53-32) on top of it. The backup ring should
be oriented with the taper pointing toward the o-ring.
8. Slide the bob shaft into the SGSM base.
9. Place the magnet cover over the magnets and screw it tight.
Bob Shaft
SGSM Base
Magnets