32
1. The test stops automatically. The alarm needs to be acknowledged and
cleared by clicking the SGSM alarm (green button) that appears in the
upper right hand corner.
Always open the pressure release valve very slowly to prevent
pulling cement into the plumbing.
2.
Unscrew and remove the high-pressure line from the cell cap and the
instrument cabinet.
Use caution. The test cell may contain a small amount of air at high
temperature and/or pressure.
3.
Unplug the thermocouple.
4. Remove the SGSM cap assembly.
5. Remove the thermocouple.
6. Lift the cell out of the heating jacket.
Operation
Stopping the Test