OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
53
Appendix
Manual Temperature
Control
The Eurotherm temperature controller allows the user to program a
temperature profile manually. The Eurotherm is in manual mode by default
but will change to auto mode when the test begins. The user must change it
to manual mode after the test has started.
This profile will be divided into at least two segments. Each segment
represents either a change in temperature or a period of time to hold the
current temperature.
The four buttons along the bottom of the display provide access to the
temperature controller parameters. To program a test:
1.
Press the “PAGE” button until the display reads “Prog List”.
2. Press the “SCROLL” key to select the parameter you want to change.
3. Press either arrow key to set the value for the parameter.
4. Repeat steps 2 and 3 until all parameters are set.
The PID and other internal settings have been optimized for this
instrument. Do not change anything other than program parameters.
For more information, refer to the Eurotherm instruction manual.
The first group of parameters will be the same for every test. Do not change
them:
Parameter
Value
Prg1 1
Hb OFF
Hb.u 0.0
Rmp.u min
Dwl.u min
Cyc.n 1
Temperature Controller
RUN/HOLD
Page Scroll