OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
55
Example 1:
Heat the sample at 2.5° per minute and stop at 150°. Hold at 150° for 180
minutes and then stop the heat.
Parameter
Value
Description
Prg 1
Hb OFF
Hb.u 0.0
Rmp.u min
Dwl.u min
Cyc.n 1
Seg
1
Segment 1
Type
rmp.r
Increase temperature at a specified rate
Tgt
150
Heat to 150°
Rate
2.5
Increase temperature at 2.5° per minute
Seg
2
Segment 2
Type
dwell
Hold on the current temperature
Dur
180
Hold for 180 minutes
Seg
3
Segment 3
Type
end
This is the last segment
End.t
sop
Stop the heat
Example 2:
Heat the sample to 200° over a period of 90 minutes. Then increase the
temperature to 300° at a rate of 3° per minute. Hold that temperature until the
unit is turned off.
Prg 1
Hb OFF
Hb.u 0.0
Rmp.u min
Dwl.u min
Cyc.n 1
Seg
1
Segment 1
Type
rmp.t
Increase temperature for a specified time
Tgt
200
Heat to 200°
Dur
90
Increase temperature for 90 minutes
Seg
2
Segment 2
Type
rmp.r
Increase temperature at a specified rate
Tgt
300
Heat to 300°
Rate
3
Increase temperature at 3° per minute
Seg
3
Segment 3
Type
end
This is the last segment
End.t
dwell
Hold at the current temperature indefinitely
Do not change
Do not change