1.
Begin by defining the first segment of the test.
a. Press the “SCROLL” key until “Seg” appears on the display.
b. Press either arrow key to select “1” (segment 1).
2.
The first parameter is “Type”.
a. Press the “SCROLL” key to select “Type”.
b. The available options are “rmp.r”, “rmp.t”, or “dwell”. Press either
arrow key to select one.
Rmp.r
programs the controller to steadily increase the temperature by
a specified rate (degrees per minute). If you choose this option, the
next parameter will be “Tgt” (target temperature) and then “Rate”
Rmp.t
increases the temperature over a specified time interval
(minutes). If you choose this option, the next parameter will be “Tgt”
(target temperature) and then “Dur” (duration in minutes).
Dwell
holds the temperature at its current set point for a specified
length of time. If you choose this option, the next parameter will be
“Dur” (duration in minutes).
c. Press the “SCROLL” key to select the remaining parameters (target
temperature, rate, or duration).
d. Press either arrow key to select the options for each parameter.
3.
Now define the second segment.
a. Press the “SCROLL” key until “Seg” appears on the display.
b. Press either arrow key to select “2” (segment 2).
4. Continue this process with each segment in the test.
5. When you reach the last segment, set the “Type” to “end”. The next
parameter will be “End.t”.
If you choose
“sop”
, the heat will be turned off and the test ended.
If you choose
“dwell”
, the heat will be held at the current temperature
indefinitely.