OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
47
10. Hold the SGSM base with the threaded end of the shaft pointed up. Place
the lower backup ring (#120-53-59) around the bottom of the Mounting
Adapter with the o-ring (#120-53-31) on top of it. The backup ring should
be oriented with the taper pointing toward the o-ring. Make sure the
backup ring and o-ring are centered around the Mounting Adapter.
11. Carefully lower the cell cap onto the shaft, making sure the backup ring
and o-ring remain in proper alignment. Screw the two pieces together
hand tight.
Retainer Ring
(#120-58-05)
O-ring
(#120-53-31)
Backup Ring
(#120-53-59)
O-ring and Backup Ring in
Proper Alignment
Magnet Cover
Top Cell Cap
Bob
SGSM Base
Bob Shaft