4. Plug the thermocouple from the cell cap to the port on the cabinet leaving
the thermocouple gland slightly loose on the cell cap.
5. Turn the water on until water comes out of the thermocouple gland and
ventilation port on the top cell cap.
a.
On the Single Cell UCA (#120-50), turn the “WATER SUPPLY” valve
to “ON”.
b.
On the Twin Cell UCA (#120-51) turn the “CELL 1 WATER” or “CELL 2
WATER” valve to “FILL”.
c.
On the Dual Cell UCA (#120-52), turn the appropriate “FILL” valve to
“ON”.
6. Once the water comes out of the ventilation port, tighten the thermo-
couple gland completely and turn the water off. This will ensure that all air
has been purged from the cell.
7. Clean and dry the top cell cap and cabinet of excess water.
8.
Carefully place the SGSM Drive Assembly on top of the test cell. Make sure
all the cables are connected properly. See page 7 for more information.