*IDN?
Command for checking load's information.
There are 3 information you can check. (Producer, Product model name, detail info).
Return value
"ODA Technologies,OPC-3010,1.0-1.0-1.0"
1
Manufacturer
2
Product model name
3
Product detail information.
1
System controller Version
2
Front panel Version
3
SCPI protocol Version
ex) *idn? return value "ODA Technologies,OEL-Series,1.0-1.0-1.0"
*SAV {1|2|3|4|5|6|7|8|10}
Command for saving data at non-volatile memory
> 1 ~ 10 Memory number
ex) *sav 2
Save data at memory 2
*RCL {1|2|3|4|5|6|7|8|10}
Command for applying all the saved settings in 『non-volatile memory』.
> 1 ~ 10 Memory number
ex) *rcl 2
Apply memory 2 data to device.
*RST
Command for initializing load's setting.
Below is the setting value after initialization.
>Voltage
0V
>Dynamic A
0
>Current
0A
>Dynamic B
0
>MODE
CC
>DUTY
50%
>Range
Low
>FREQ
0.0001
>SLEW
LEVEL 1
>Input Mode OFF
- 88 -