▌Press A/Calibration key to save the setting value.
"ADC-DATA VALUE 0980" will be displayed.
ps. Display value can be different.
Go back to before Cal status.
▌
Press calibration key in
"4
.CC-Range:LOW_LOW
" status.
"CV-HIGH 000.4A" will be displayed.
ps. Display value can be different.
Wait for 5 min
▌Wait until DVM Meter shows stable voltage.
▌Calculate current value.
For example, if resistance value is 0.01
Ω
, checked voltage is 61.5mV.
Current value will be 61.5A.
▌Set calculated current value to the power supply.
▌Press A/Calibration key to save the setting data.
"ADC-DATA VALUE 0016" will be
ps. Display value can be different.
Go back to before Cal status.
- 57 -
A/Cal