▌After setting, press A/Calibration key to save the setting value.
"ADC-DATA VALUE 0970" will be displayed.
ps. Display value can be different.
Go back to before Cal status.
▌
Press Calibration key in
"2
.CC-Range:HIGH_LOW
" status.
"CC-HIGH_LOW 000.4A" will be displayed.
ps. Display value can be different.
Wait for 5 min
▌Wait until DVM Meter shows stable voltage.
▌Calculate current value.
For example, if resistance value is 0.01
Ω
, checked voltage is 61.5mV.
Current value will be 61.5A.
▌Set calculated current value to the power supply.
▌After setting, press A/Calibration Key to save the setting value.
"ADC-DATA VALUE 0016" will be displayed.
ps. Display value can be different.
Go back to before Cal status.
▌
Press Calibration key in
"3
.CC-Range:LOW_HIGH"
status.
"CC-LOW_HIGH 006.2A" will be displayed.
ps. Display value can be different.
Wait for 5 min
▌Wait until DVM Meter shows stable voltage.
▌Calculate current value.
For example, if resistance value is 0.01
Ω
, checked voltage is 61.5mV.
Current value will be 61.5A.
▌Set calculated current value to the power supply.
- 56 -
A/Cal