75
Norsonic Nor140
Instruction Manual
Two printout examples, a profile and a sound power meas-
urement output
To set which functions to print out:
1
Press
SETUP >1 >9 >1
(Depend on options).
2
Use the cursor keys to navigate and the
INC
and
DEC
keys to set the functions to be printed. A “
1” means that
the function will appear in the printout.
3
Units equipped with the enhanced profile extension
allow you to select which functions to log as level vs.
time. Once you’ve set up this table and pressed enter
to put changes into effect, the instrument will prompt
you to decide whether this setup shall be copied to the
Prnt par. menu as well, for convenience. Note that this
applies to profile printouts only.
To start a printout:
1
Set up everything as required.
2
Press the
key.
For level and frequency analysis measurements the
entire data set will be printed out. For profile measure-
ments the printout will contain the periods from the
present cursor position till the end of the time record.
How to print the level, the profile and the frequen-
cy spectrum:
• You decide what to print by what is on the display
you press the print key. This means that to print
the spectrum, you must display the spectrum on
the screen, to print the profile you must display the
profile and so on.