159
Norsonic Nor140
Instruction Manual
159
Norsonic Nor140
Instruction Manual
12 requires all the options 1, 3, 9, 10 and 11 to be in-
stalled in the Nor140 unit.
Swept-Sine measurement of sound level
Upon connecting the Nor140 to the PC running the
NorBuild software package with the CtrlBuild remote
feature, the selection of Swept-Sine technique is made
in the Measurement Mode menu of the software. Click
on the point for “Level Swept-Sine”.
The different choices for the actual Swept-Sine level
measurement are then displayed in the Measurement
Setup menu.
In the Measurement menu, there are two settings that
differ from the normal operation of the NorBuild soft-
ware package.
“Sweep duration” is used for selecting the speed of
the sinus sweep used in the measurement. This can
be set to “Short”, Medium” or “Long” which makes the
duration lasting approx. 1, 5 or 11 minutes respectively.
The longer duration, the better S/N ratio is achieved.
“Required S/N-ratio” is setting the acceptance ratio
for the measured results. Should any frequency band
not give the required S/N-ratio, the value will be shown
with a “?” in the result table.
The other settings are described in the instruction
manual delivered with the NorBuild software package.
Swept-Sine measurement of reverberation time
Click on the point for “Reverberation Swept-Sine” in
the Measurement Mode menu, and follow the same
selection possibilities as described above for the level
swept-sine measurements.