178
Chapter 28
Audiometer calibration
Audiometer calibration
Audiometer calibration is made easily with a Nor140
sound level meter (software option; audiometer cali-
bration and 1/3-octave frequency analysis is required).
Frequency, Level and Distortion is measured and pre-
sented after each single measurement.
To activate the audiometer calibration software in
Nor140, press
SETUP > 1 > 9
Misc .par, choose
5
Audiometer and toggle audiometer calibration function
ON/OFF with
INC/DEC
keys. Press
ENTER
repeatedly
to confirm and leave menus.
Connect and calibrate Nor140 to an artificial ear, for
example. Gras 43AA. Play the desired tone (frequency
and level) of the audiometer connected to the artificial
ear.
Press
START
on Nor140, measurement time is by default
2 sec but can be changed by the user. Nor140 presents
after a single measurement, the precise frequency, level
and distortion on the display. The measurement can be
saved and exported to NorXfer PC software.