187
Norsonic Nor140
Instruction Manual
187
Activating the high level range:
• Press
SETUP
>
5
(Lvl.range) to gain access to the level
range setting menu. Navigate in the menu as usual.
Preamplifi er attenuation
The instrument has the ability to correct for the attenu-
ation in the preamplifi er. Typical value of the attenua-
tion is 0.7 dB. The correction can be set in the range
0.0 to 9.9 dB. The correction can be switched on/off to
facilitate applications of other transducers without pre-
amplifi er. The correction is automatically switched off if
IEPE-type of preamplifi er is selected.
Activating the preamplifi er attenuation:
• To activate the preamplifi er attenuation press
Press
SETUP
>
1
(Instr.) >
4
(Input) >
2
(Correc-
tions)
to gain access to the Corrections menu. Navi-
gate in the menu as usual and activate the correc-
tion parameter Preamp by means of the inc and
dec keys. Do the same to deactivate. Preamplifi er
attenuation activated is indicated by a
G
(for Gain)
in the lower line of the display
Setting the amount of attenuation:
• In the Correction menu, press
1
(Corr.par) to gain
access to the correction parameter setup menu.
Press
2
to gain access to the attenuation setting
itself. Never change this setting unless you know
what you’re doing! Use the numerical keypad to set
the attenuation value. Press
ENTER
twice to leave
the menu. See Fig. on the previous page for menu
details