Storing the result
The measured spectrum may be stored in the instru-
ment and later retrieved or exported to a PC by using
the NorXfer program. The storing is as simple as in the
normal mode of operation. A file number is automati-
cally generated. In the file list the file number is suc-
ceeded with the letter “f” indicating a FFT-result. The
storing may be manual or automatic as in the normal
mode of operation.
Specification for FFT
Sampling frequency: 24 kHz
FFT size:
16384 samples (0.682666 sec)
Frequency lines:
8192 frequency lines
6553 will be available
(after LP anti-aliasing filter)
Anti-aliasing filter:
Pass-band ripple: 0.03 dB
Stop band attenuation: 100 dB
Frequency range:
0 … 9600 Hz (-1 dB point)
Frequency resolution:
Δ
f
= 1/T = 1.46… Hz (24000/2
14
)
Window function:
Hanning
FFT function:
Auto spectrum
Spectrum unit:
dB and EU (Engineering units)
Averaging:
Linear power averaging
Number of averages: 1 to 1028571
Trigger:
Manual, External, Clock,
Level above (Option)
Input source:
Microphone socket
Dynamic range:
120 dB, 80 dB displayed.
Zoom:
Display function (C= 1, 2, 4, 8, ... 64)
If the display is compressed, more spectral
lines are displayed as one line. The cursor
value will show the maximum value for all
the lines represented by the cursor posi-
tion. In order to read the values for each spectral
line, you have to zoom until the compression factor
is one (C = 1)!