Chapter 29
Technical Specifications
Measurement range
Total range for measurement of A-weighted levels
The linear operating range is identical to the total
range.
Frequency
31.5 Hz 1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
98 dB 137 dB 138 dB 136 dB 133 dB
Lower level
24 dB 24 dB 24 dB 24 dB
24 dB
Ref level test
94 dB 114 dB 114 dB 114 dB 114 dB
The primary indicator range for compliance with IEC
60651 type 1 is 24 dB to 117 dB. For compliance with
IEC 60804 type 1, the linearity range is 24 to 137 dB, and
the pulse range 24 dB to 140 dB, respectively
Total range for measurement of C-weighted levels
The linear operating range is identical to the total range.
Frequency
31.5 Hz 1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
134 dB 137 dB 136 dB 134 dB 131 dB
Lower level
30 dB 30 dB 30 dB 30 dB
30 dB
Ref level test
114 dB 114 dB 114 dB 114 dB 114 dB
Total range for measurement of Z-normal -
weighted levels
The linear operating range is identical to the total range.
Frequency
31.5 Hz 1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
137 dB 137 dB 137 dB 137 dB 137 dB
Lower level
40 dB 40 dB 40 dB 40 dB
40 dB
Ref level test
114 dB 114 dB 114 dB 114 dB 114 dB
Total range for measurement of Z-wide -weighted
levels
The linear operating range is identical to the total range.
Frequency 3.15 Hz 31.5 Hz 1 kHz
4 kHz 8 kHz 12.5 kHz
Upper level
137 dB 137 dB 137 dB 137 dB 137 dB 137 dB
Lower level
50 dB
50 dB 50 dB 50 dB 50 dB 50 dB
Ref level test 114 dB
114 dB 114 dB 114 dB 114 dB 114 dB
Measurement range for C-weighted peak levels
Frequency
31.5 Hz 1 kHz
4 kHz
8 kHz 12.5 kHz
Upper level
137 dB 140 dB 139 dB 137 dB 134 dB
Lower level
45 dB 45 dB 45 dB 45 dB
45 dB
Ref level test
114 dB 114 dB 114 dB 114 dB 114 dB
Power supply
Battery:
4 cells, IEC LR6, AA-sized, Alkaline bat-
teries are recommended (e.g. Duracell Ultra
M3). AA-sized NiCd or NiMH rechargeable bat-
teries may be used, but must be charged outside
the instrument. Battery voltage and time on bat-
tery since last change of batteries are indicated.
Typical battery life time
(Duracell Ultra M3)
:
8 hours.
External
DC
:
11 – 16 volt. Power consumption ap-
proximately 1.2 watt dependent on selected modes
of operation. External DC source should have source-
impedance less than 1 ohm and be able to supply at
least 300 mA. The mains adaptor Nor340 is recom-
mended for use with the instrument.
If the external supply falls below 9 V, the instrument
will use the internal batteries if available. If the instru-
ment has switched off due to loss of power or insuf-
ficient supply voltage, the instrument will automatically
switch on and resume normal operation after reapply-
ing the external DC supply.