75
5.2.74 Error 10002 (Emergency button is pressed.)
5.2.75 Error 10003 (Too thin lens)
Replace the lens with a thicker one.
5.2.76 Error 10007 (No right-lens data)
Load the right lens shape data.
5.2.77 Error 10008 (No left-lens data)
Load the left lens shape data.
5.2.78 Error 10010 (Too thick lens)
Replace the lens with a thinner one.
5.2.79 Error 10012 (Too large lens shape)
Reduce the lens size.
5.2.80 Error 10014 (Bevel position comes off.)
Adjust the bevel position so that it does not come off the lens edge.
Is the emergency stop button pressed?
Yes
Release the emergency stop button.
No
Are there any breaks in the cable ASSY (40274-
CA31)?
Yes
Replace the cable ASSY (40274-CA31)
.
No
Replace the I/O board (40274-BA06)
.
Summary of Contents for SE-9090 Supra
Page 1: ...SYSTEM EDGER SE 9090 Supra SERVICE MANUAL September 2012 Pages in total 350 LLE7D RDA002B...
Page 12: ...LLE7D RDA002B X This page is intentionally left blank...
Page 14: ...LLE7D RDA002B 2 This page is intentionally left blank...
Page 22: ...LLE7D RDA002B 10 This page is intentionally left blank...
Page 24: ...LLE7D RDA002B 12 This page is intentionally left blank...
Page 104: ...LLE7D RDA002B 92 This page is intentionally left blank...
Page 310: ...LLE7D RDA002B 298 9 3 Wiring Diagram...
Page 311: ...LLE7D RDA002B 299 9 4 Connector Cable...
Page 312: ...LLE7D RDA002B 300...
Page 313: ...LLE7D RDA002B 301 9 5 Layout Drawing 9 5 1 ASSYs 40274 2000...
Page 314: ...LLE7D RDA002B 302 40274 3000 40274 5000 40274 6000...
Page 315: ...LLE7D RDA002B 303...
Page 316: ...LLE7D RDA002B 304 40274 4000 40274 8000...
Page 317: ...LLE7D RDA002B 305 40274 7100 40274 7200 40273 7200 2 2 5 2 5 5 2 2 2 5...
Page 318: ...LLE7D RDA002B 306 9 5 2 Boards S N 700019 and earlier...
Page 319: ...LLE7D RDA002B 307 S N 700020 and later...
Page 320: ...LLE7D RDA002B 308 9 5 3 Cable ASSYs...
Page 321: ...LLE7D RDA002B 309 9 5 4 Other electrical components...
Page 322: ...LLE7D RDA002B 310...
Page 350: ...LLE7D RDA002B 338 This page is intentionally left blank...