115
2 ) Loosen the machine screws (40273-
M405).
3 ) Unscrew FC4 × 8 (n = 4) to loosen
the right diaphragm holder (40273-
M406).
4 ) Turn the grooving ASSY (40274-
7210) in the direction of the arrow
completely.
5 ) Unscrew SB6 × 20 (n = 3) at the
back of the diaphragm (40274-
M715).
/
/
(%
/
5$
/
M715