176
8.4 Chuck Pressure Measurement
2 . Press the Parameter setting button, then
select Tab 2.
3 . Confirm that the Chuck pressure parame-
ter is set to “8”.
4 . Display the hardware test screen
.
5 . Press Chuck [Pow
↑
] to set the chuck
operation.
6 . Press Chuck [Main] to close the chuck.
Press Chuck [Open] to open the chuck.
7 . Open and close the chuck 5 to 10 times,
then attach the meter of the chuck pres-
sure measurement jig (40391-0100) to the
lower shaft with the chuck open.
8 . Close the chuck and confirm that the meter reads 60 ±10 kgf.
9 . If the measured value is not within the range, change the Chuck pressure parameter value
so that the specified range is met.
Set the same value as con-
firmed in Step 3.
Summary of Contents for SE-9090 Supra
Page 1: ...SYSTEM EDGER SE 9090 Supra SERVICE MANUAL September 2012 Pages in total 350 LLE7D RDA002B...
Page 12: ...LLE7D RDA002B X This page is intentionally left blank...
Page 14: ...LLE7D RDA002B 2 This page is intentionally left blank...
Page 22: ...LLE7D RDA002B 10 This page is intentionally left blank...
Page 24: ...LLE7D RDA002B 12 This page is intentionally left blank...
Page 104: ...LLE7D RDA002B 92 This page is intentionally left blank...
Page 310: ...LLE7D RDA002B 298 9 3 Wiring Diagram...
Page 311: ...LLE7D RDA002B 299 9 4 Connector Cable...
Page 312: ...LLE7D RDA002B 300...
Page 313: ...LLE7D RDA002B 301 9 5 Layout Drawing 9 5 1 ASSYs 40274 2000...
Page 314: ...LLE7D RDA002B 302 40274 3000 40274 5000 40274 6000...
Page 315: ...LLE7D RDA002B 303...
Page 316: ...LLE7D RDA002B 304 40274 4000 40274 8000...
Page 317: ...LLE7D RDA002B 305 40274 7100 40274 7200 40273 7200 2 2 5 2 5 5 2 2 2 5...
Page 318: ...LLE7D RDA002B 306 9 5 2 Boards S N 700019 and earlier...
Page 319: ...LLE7D RDA002B 307 S N 700020 and later...
Page 320: ...LLE7D RDA002B 308 9 5 3 Cable ASSYs...
Page 321: ...LLE7D RDA002B 309 9 5 4 Other electrical components...
Page 322: ...LLE7D RDA002B 310...
Page 350: ...LLE7D RDA002B 338 This page is intentionally left blank...