5.2.62 Error 2004 (Grooving motor error)
5.2.63 Error 2101 (RH motor communication error)
Are there any breaks in the cable ASSY (40274-
CA27)?
Yes
Replace the cable ASSY (40274-CA27)
.
No
Replace the grooving ASSY (40274-721S)
.
Is the problem resolved?
Yes
End
No
Replace the I/O board (40274-BA06)
.
Are there any breaks in the cable ASSY (40273-
CA18) for R.H.PM?
Yes
Replace the cable ASSY (40273-CA18) for R.H.PM
(see 9.3 [p298] and 9.4 [p299])
No
Replace the closed loop motor driver (40274-E019)
for R.H.PM
.
Is the problem resolved?
Yes
End
No
Replace the I/O board (40274-BA06)
.
Summary of Contents for SE-9090 Supra
Page 1: ...SYSTEM EDGER SE 9090 Supra SERVICE MANUAL September 2012 Pages in total 350 LLE7D RDA002B...
Page 12: ...LLE7D RDA002B X This page is intentionally left blank...
Page 14: ...LLE7D RDA002B 2 This page is intentionally left blank...
Page 22: ...LLE7D RDA002B 10 This page is intentionally left blank...
Page 24: ...LLE7D RDA002B 12 This page is intentionally left blank...
Page 104: ...LLE7D RDA002B 92 This page is intentionally left blank...
Page 310: ...LLE7D RDA002B 298 9 3 Wiring Diagram...
Page 311: ...LLE7D RDA002B 299 9 4 Connector Cable...
Page 312: ...LLE7D RDA002B 300...
Page 313: ...LLE7D RDA002B 301 9 5 Layout Drawing 9 5 1 ASSYs 40274 2000...
Page 314: ...LLE7D RDA002B 302 40274 3000 40274 5000 40274 6000...
Page 315: ...LLE7D RDA002B 303...
Page 316: ...LLE7D RDA002B 304 40274 4000 40274 8000...
Page 317: ...LLE7D RDA002B 305 40274 7100 40274 7200 40273 7200 2 2 5 2 5 5 2 2 2 5...
Page 318: ...LLE7D RDA002B 306 9 5 2 Boards S N 700019 and earlier...
Page 319: ...LLE7D RDA002B 307 S N 700020 and later...
Page 320: ...LLE7D RDA002B 308 9 5 3 Cable ASSYs...
Page 321: ...LLE7D RDA002B 309 9 5 4 Other electrical components...
Page 322: ...LLE7D RDA002B 310...
Page 350: ...LLE7D RDA002B 338 This page is intentionally left blank...