145
OPERATION FOR WHEN PERIPHERAL DEVICES ARE CONNECTED
:
Connection of PC
3
Connection
1
Connect the device to the network.
Connect the device to the network via LAN to which the PC at the receiving end is con-
nected.
2
Turn on power to the PC, then to the device.
3
Set parameters for the device and the PC after obtaining permission from the network
administrator of the facility.
For details of the parameter setting, see “
Entrust the connection to service personnel of NIDEK or your authorized distributor.
4
Check the connection between the device and the PC.
There are two methods to check the connection.
1) After inputting the IP address of the PC in the Network Info. screen, transmit the ping
command by pressing the Transmit button. (“
2) When the device is connected to the NAVIS-EX, input the IP address and such in the
“External settings—Transfer (NAVIS-EX)” screen, then press the “Test Connection”. (“
• When the device is disconnected with NAVIS-EX, the following message appears:
NAVIS-EX has been disconnected.
Would you like to reconnect to NAVIS-EX?
When connecting the device with NAVIS-EX again, check the LAN cable and the receiving
PC to make sure that the device and the PC are properly connected, then press the Yes
button on the screen.
AFC-330
Receiving PC
LAN cable
Network hub (HUB)