60
OPERATING PROCEDURE
:
Image Capture (Normal Photography Mode [Auto shot, No blink])
5
Perform alignment using the joystick so that the patient’s anterior eye is displayed on the
LCD touch-screen.
6
Make the (eight) cornea-reflected spots appear.
When the eight cornea-reflected spots are detected, the alignment automatically proceeds
as described below.
* Here, the internal fixation lamp is in the center of the “STD.” setting.
To change the internal fixation lamp position, see “2.8.1 Fixation lamp mode” (Page 101).
* These cornea-reflected spots may not be detected depending on the condition of the
patient’s eye, and the auto tracking may not be activated. (See “ Auto tracking cannot be
executed.” (Page 148) in “4.1 Troubleshooting” (Page 147).) In such a case, alignment
needs to be performed manually. (See Steps 3 to 5 in “2.3.2 Image Capture without auto-
matic functions” (Page 65).)
Forehead rest
Eye level marker
Chinrest
Chinrest up/down buttons