SPECIFICATIONS AND ACCESSORIES
5.1
Classifications
[Form of protection against electrical shock] Class I
The AFC-330 is classified as a Class I device. A Class I device is a device in which protec-
tion against electric shock does not rely on basic insulation only, but which includes addi-
tional safety precaution in such a way that means are provided for the connection of
accessible conductive parts to the protective ground conductor in the fixed wiring of the
installation so that they cannot become live in the event of a failure of the basic insulation.
[Compliance with Electromagnetic Compatibility Directive]
The AFC-330 complies with the EMC standard, IEC 60601-1-2: 2007
[Degree of protection against harmful ingress of water or particulate matter] IPX0
*1
The AFC-330 provides no protection against ingress of water or particulate matter with
harmful effects. Avoid splashing water or other liquids on or near the device.
[Degree of suitability for use in an oxygen rich environment]
The AFC-330 is not intended for use in an oxygen rich environment.
[Method(s) of sterilization recommended by the manufacturer]
The AFC-330 does not have any part to be sterilized.
[Mode of operation]
The AFC-330 is classified as a continuous operating device.
[Mode of transport]
The AFC-330 is classified as a stationary device.
*1.
In accordance with IEC 60529