13
BEFORE USE
:
Screen Configuration
1
11. Vertical movement limit indicator
Indicates that the image capturing unit is at the limit of the vertical movement range in the anterior eye and
fundus observation screens.
When this icon is displayed in the upper half of the screen, the image capturing unit is at the upper
limit of the vertical movement range. Lower the image capturing unit.
When this icon is displayed in the lower half of the screen, the image capturing unit is at the lower limit
of the vertical movement range. Raise the image capturing unit.
12. Left and right movement limit indicator
Indicates that the image capturing unit is at the limit of the left or right movement range in the anterior eye or
fundus observation screens.
When this icon is displayed on the screen, the image capturing unit is at the limit of the left
movement range. Tilt the joystick a little to the right.
When this icon is displayed on the screen, the image capturing unit is at the limit of the right
movement range. Tilt the joystick a little to the left.
13. Required pupil diameter detection mark
,
Indicates whether or not the patient’s eyelid is covering their eye and whether or not their pupil size satisfies
the required pupil diameter (or the minimum pupil diameter in Small pupil photography mode).
: Indicates that the patient's eyelid is not covering their eye, and that their pupil size satisfies the
required pupil diameter (or the minimum pupil diameter in Small pupil photography mode). When this
mark is displayed, image capture can be executed.
(black and white): Indicates that the patient’s eyelid is covering the patient’s eye, or that their eye is not
detected.
(color): Indicates that the patient’s eyelid is detected, the pupil size does not satisfy the required pupil
diameter (or the minimum pupil diameter in Small pupil photography mode) in the anterior eye
or fundus observation screen, or that extraneous light or such (noise) is detected.
This mark is not displayed in Anterior eye photography mode.
14. Low-light photography toggle button
Used to lower the flash intensity for image capture by the specific amount from the current flash intensity.
For details of this button, see “2.5.5 Low-light photography mode” (Page 79).
This button is displayed when the Low-Light setting is ON in the Photography Modes screen (not displayed
when the setting is OFF).
• When the auto tracking is enabled, rotating the joystick to raise the image capturing unit
may not make the vertical movement limit indicator icon disappear. In such a case, raise the
chinrest.