117
OPERATING PROCEDURE
:
Control Menu
2
4
When selection of the internal fixation lamp positions is finished, press the Done button.
The Done button changes to the Select button, and the Menu button becomes enabled.
To restore the default setting, press the Default button at the bottom left of the screen.
5
When the setting is complete, press the Menu button.
The Control menu 1/2 screen is displayed.
CAM1, CAM2, CAM3
1
Press any of “CAM1”, “CAM2”, and
“CAM3” in “Peripheral settings”.
The screen for setting the parame-
ters of the selected camera setting is
displayed.
The procedure is explained with
“CAM1”.
2
Set the parameters in the CAM1 1/2
screen.
* Default settings are underlined.
[Control menu 1/2]
Advance button
Return button
Increase button
Reduce button
Item
Options
ISO speed
100 / 200 / 400
The camera speed can be changed. Compared with ISO 100, images of
the same brightness can be captured with half the intensity using ISO 200
and 1/4 of the intensity using ISO 400.
However, the increasing the ISO speed increases the noise in the image.
In Anterior eye photography mode, the parameter is fixed to ISO 100.