58
OPERATING PROCEDURE
:
Image Capture (Normal Photography Mode [Auto shot, No blink])
2.3
Image Capture (Normal Photography Mode [Auto shot, No blink])
Image capture is performed after selecting the desired patient from the patient information register
screen.
In non-mydriatic photography, the patient’s pupil needs to be dilated in a darkened place (a light level
at which a newspaper can barely be read).
The operating procedure below is explained in Normal photography mode using the auto shot without
the blink detector and without auto switching to Small pupil photography mode that is the default set-
ting.
2.3.1
Image Capture with auto shot
Image capture with Auto
selected with the Auto/Manual toggle button
1
Press the patient list button in the anterior eye observation screen.
The patient information register screen appears.
2
Press the desired patient for whom images are to be captured.
The row of the selected patient is highlighted in orange.
Patient list button
Anterior eye observation screen
37 $EFG()JK
)HPDOH \HDUPRQWK
Patient information register screen
Patient register button
37 $EFG()JK
)HPDOH \HDUPRQWK
Patient register button