** Enter Setup
27
Exit Setup
Timeout between Decodes
Timeout between Decodes:
This parameter sets the time period between the end of one decode
session and the start of next session. It is programmable in 500ms increments from 0ms to 7,500ms. The
default timeout is 0ms. To program this parameter, scan the
Timeout between Decodes
barcode, the
numeric barcodes and the
Save
barcode.
Timeout between Decodes
Example: Set the timeout between decodes to 1,000ms
1. Scan the
Enter Setup
barcode.
2. Scan the
Timeout between Decodes
barcode.
3. Scan the numeric barcodes “1”, “0”, “0” and “0” from the “Digit Barcodes” section in Appendix.
4. Scan the
Save
barcode from the “Save/Cancel Barcodes” section in Appendix.
5. Scan the
Exit Setup
barcode.