![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339093.webp)
Patient List on Programmer and the PDMS on a PC
89
RNS
®
System Programming Manual
•
When accessing the PDMS via a web browser (not via the tablet), you can define or edit
and save a detection set to the PDMS, and later download the detection set to the
programmer. Having saved a detection set to the programmer, you can program the
neurostimulator while the programmer is offline, as is required, for example, if an
operating room lacks network connectivity. Note that these instructions apply whether the
target programmer is a model PGM-300 (laptop) or model 5000 (RNS
®
Tablet).
•
S
TIMULATION
screen: This screen enables you only to review the patient’s history of
stimulation settings. You can adjust stimulation settings only using the tablet during an
interactive programming session. See
on page
Figure 75 Stimulation screen through the patient list—review settings history only