Tablet Instructions
45
RNS
®
System Programming Manual
In future visits, after collecting additional ECoGs, you then, in
Phase 3
, iterate to improve detection (and,
once responsive therapy has been enabled, to improve clinical response) by making small, incremental
changes to detection parameters with the sliders or
M
ORE
C
ONTROLS
.
Note: Typically, physicians repeat earlier phases (Phase 1 or Phase 2) only if they observe a new
type of ECoG activity of interest that requires an additional detector; or if a good clinical response
has not been obtained and they wish to start over.
Phase 1: Select a Default Detection Set upon First Use
Upon first use of the tablet for each patient, the patient has no saved ECoGs from which to select a
pattern for detection, and so you must select one of three default detection sets. This is usually
configured in the operating room immediately after implantation. The default sets are used to gather
an initial set of ECoGs for a patient, which you can then use in a subsequent visit to define a pattern
detection set specific to the patient. If you have already performed these steps once, proceed to the
next section,
Phase 2: Define a Pattern Detection Set
1. Select
N
EUROSTIM
S
ETTINGS
from the
H
OME
screen.