![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339053.webp)
Tablet Instructions
49
RNS
®
System Programming Manual
the ECoGs shown using the
T
RIGGERS
and
C
ATEGORIES
menus at left center, and/or by date
range using the
F
ROM
and
T
O
fields at middle right.
•
The
S
IMULATE
button at upper right (next to
A
DVANCED
) is switched
O
N
(blue) by default,
which enables simulations of detections to be shown on all ECoGs shown on this screen.
3. To define or change a detection set, select the desired detection set in the
N
AME
field at
upper left.
•
The detection set may be one of the three default detection sets to be used as a starting point
for configuration; or it may be a detection set created and saved previously.
•
If you don’t want to modify the detection set you chose, you can program the detection set
right away: Select
R
EVIEW
&
P
ROGRAM
at upper right and, while holding the wand within
approximately 1 inch of the neurostimulator, follow the prompted steps to
C
ONFIRM
P
ROGRAMMING
and
C
ONTINUE
sending the new settings to the neurostimulator.
4. Use the
D
EFINE
P
ATTERNS
dialog to define or modify a detection set. There are two routes to
the
D
EFINE
P
ATTERNS
dialog.
•
Select one of the two
P
ATTERN
tiles at the top to go there directly, or
•
Select one of the ECoG thumbnails at the bottom to open its ECoG dialog, which is named by
its date and trigger type, and then select one of the two
M
ODIFY
buttons (
A1
&
A2
,
B1
&
B2
)
at upper right. (You can also use the ECoG dialog to assign the ECoG to a category if
desired.)
If You Selected a New ECoG First
5. If you selected a new ECoG first and selected a
M
ODIFY
button, a
S
ELECT
C
HANNEL
dialog
opens. Touch within the channel that has a pattern to detect. Then the
D
EFINE
P
ATTERNS
dialog opens.
Summary of Contents for PDMS 4340